72
Division
Disassembling the main body
Serial No.
Section
Disassembling the main body
Descriptions
Removing the upper parts of the main
body and the humidity chamber
Tools and jigs to be used
No
Part name
Qty
No
Disassembling procedure
Remarks
1
Main body case 2200
1
1
Open the humidity chamber cover
of
①
, and remove
⑥
.
2
Top board (12
°
tilting)
1
2
Open the hood, and remove
⑤
which is placed on
②
.
3
Hood (slope type DA)
1
3
Remove
④
from
③
. Remove the
cables of
④
from J1 and J2 on the
relay box of
①
.
4
Sensor module
1
5
Weight monitor mattress platform
1
4
Remove
③
from the hinges (right
and left) of
①
by pulling it out.
6
Humidity chamber
1
5
Remove
②
and
⑦
from the top of
①
.
7
Casing
1
6
8
7
9
8
10
9
Summary of Contents for V-2200
Page 1: ......
Page 2: ......
Page 4: ......
Page 14: ...12 4 2 Display Control Panel...
Page 187: ...185 Available only for units destined for the US...
Page 189: ...187 C60F0263 Available only for units with serial No 1680000 or higher...
Page 190: ...188 C60F1003 Available only for units destined for the US...
Page 191: ...189 9 2 HL stand...
Page 192: ...190...
Page 193: ...191 10 Replacement Parts List Replacement Parts List...
Page 194: ...192 10 1 Pole Arm Upper Part of the Main Body Accessories and Humidity Chamber Section Fig 1...
Page 196: ...194 10 2 Hood Section Fig 2...
Page 198: ...196 10 3 Main Body Fig 3...
Page 202: ...200 10 5 HL Stand Fig 5...
Page 211: ......
Page 212: ...C60S220 201109...