170
Division
Product inspection
Serial No.
Section
Humidity
Tools and jigs to be used
No Inspection
procedure
Remarks
1
*
When inspecting humidity, be sure to remove the oxygen suction hose and
fit the rubber stopper securely.
2
Set the incubator air temperature to 32.0
℃
and stabilize the incubator.
3
Press the relative humidity selector switch, and set the relative humidity to
90%.
After 40 minutes, check that the reading on the relative humidity display is
stable at 90
±
5%.
Relative humidity
selector switch
Setting switch