22
(2) Inspection of basic functions
①
Inspection of the basic functions of the incubator
(Preparation 1)
1. Connect the yellow skin temperature probe to
the skin temperature probe connecting port 1
on the sensor module.
2. Open the humidity chamber lever, pull down
the lever toward you, and take out the
humidity chamber. Take it out only after the
humidity chamber and the water inside have
cooled down sufficiently to avoid scalding
yourself. The humidity chamber is so heavy
that you should hold it with both your hands.
(Checking the power-on procedure)
Check:
(1) Does a numerical value appear on the
incubator air temperature display?
(2) Does a numerical value appear on the skin
temperature 1 display?
(3) Does a numerical value appear on the relative
humidity display?
(4) Does a humidity chamber off alarm appear?
(5) Are all the other alarms off?
∗
) Numerical values here shall include “LOW” and
“HIGH.”
3. Operate the power switch to activate the unit.
Summary of Contents for V-2200
Page 1: ......
Page 2: ......
Page 4: ......
Page 14: ...12 4 2 Display Control Panel...
Page 187: ...185 Available only for units destined for the US...
Page 189: ...187 C60F0263 Available only for units with serial No 1680000 or higher...
Page 190: ...188 C60F1003 Available only for units destined for the US...
Page 191: ...189 9 2 HL stand...
Page 192: ...190...
Page 193: ...191 10 Replacement Parts List Replacement Parts List...
Page 194: ...192 10 1 Pole Arm Upper Part of the Main Body Accessories and Humidity Chamber Section Fig 1...
Page 196: ...194 10 2 Hood Section Fig 2...
Page 198: ...196 10 3 Main Body Fig 3...
Page 202: ...200 10 5 HL Stand Fig 5...
Page 211: ......
Page 212: ...C60S220 201109...