28
(6)
If calibration is not completed successfully, try to
calibrate again as instructed below.
■
If you cannot proceed to the 5kg calibration procedure:
If the message “OUTSIDE 0kg CALIBRATION
RANGE” appears during calibration and the weight
monitor does not proceed to the 5kg calibration
procedure, press the Zero Switch, and make sure
that nothing is placed on the mattress platform and
that no cables are caught in the mattress platform,
etc. Then press the Mode Switch to perform the
0kg calibration procedure again.
■
If the message “OUTSIDE 5kg CALIBRATION RANGE” appears:
Press the Zero switch, and make sure that nothing
but the 5kg weight is placed on the mattress
platform and that no cables are caught in the
mattress platform, etc. Then press the Mode
Switch to perform the 5kg calibration procedure
again.
■
If the message “5kg SPAN IS OUTSIDE CALIBRATION RANGE” appears:
Press the Zero Switch. Remove the reference
weight from the mattress platform, and make sure
that nothing is placed on the mattress platform and
that no cables are caught in the mattress platform,
etc. Then press the Mode Switch to perform the 0kg
calibration procedure and the 5kg calibration
procedure again.
Summary of Contents for V-2200
Page 1: ......
Page 2: ......
Page 4: ......
Page 14: ...12 4 2 Display Control Panel...
Page 187: ...185 Available only for units destined for the US...
Page 189: ...187 C60F0263 Available only for units with serial No 1680000 or higher...
Page 190: ...188 C60F1003 Available only for units destined for the US...
Page 191: ...189 9 2 HL stand...
Page 192: ...190...
Page 193: ...191 10 Replacement Parts List Replacement Parts List...
Page 194: ...192 10 1 Pole Arm Upper Part of the Main Body Accessories and Humidity Chamber Section Fig 1...
Page 196: ...194 10 2 Hood Section Fig 2...
Page 198: ...196 10 3 Main Body Fig 3...
Page 202: ...200 10 5 HL Stand Fig 5...
Page 211: ......
Page 212: ...C60S220 201109...