![Atom Medical Corporation V-2200 Service Manual And Replacement Parts Download Page 52](http://html1.mh-extra.com/html/atom-medical-corporation/v-2200/v-2200_service-manual-and-replacement-parts_3004438052.webp)
50
6-3-2. Parts replacement record screen
(1) Description of the screen
①
The left portion of the screen shows the items of inspection and replacement parts.
②
The table on the right portion of the screen shows the details of the record of the item (the
kind of inspection or the replacement part) highlighted by the cursor.
③
The area above the table on the right portion of the screen shows the total length of time for
which the incubator has been in operation since shipment.
(2) How to record inspection and replacement
On the parts replacement record screen, you can record the date of replacement and the total
operating time of the incubator at the time when the replacement is recorded.
①
Every time the “brightness” switch is pressed, the cursor will move from one item to another
in the order of “Three-month inspection”
⇒
“Periodical inspection”
⇒
“Filter”
・・・・⇒
“Primary battery for Display”
⇒
“Three-month Inspection.”
②
Move the cursor to the item you want to record.
③
Press the “record” switch.
④
The latest recording date will be displayed at the top of the table and recorded.
⑤
Up to eight records can be displayed.
⑥
If more than eight records have been recorded, the oldest record will be erased and the eight
latest records will be displayed.
(3) How to erase a record
You can erase the records of inspection and parts replacement one by one beginning with the
latest record.
①
Move the cursor to the item you want to erase.
②
Press the Zero switch, and one latest record will be erased.
(4) How to reset the operating time
You can reset the operating time to “0” hr if you want to or when the main board has been
replaced.
Summary of Contents for V-2200
Page 1: ......
Page 2: ......
Page 4: ......
Page 14: ...12 4 2 Display Control Panel...
Page 187: ...185 Available only for units destined for the US...
Page 189: ...187 C60F0263 Available only for units with serial No 1680000 or higher...
Page 190: ...188 C60F1003 Available only for units destined for the US...
Page 191: ...189 9 2 HL stand...
Page 192: ...190...
Page 193: ...191 10 Replacement Parts List Replacement Parts List...
Page 194: ...192 10 1 Pole Arm Upper Part of the Main Body Accessories and Humidity Chamber Section Fig 1...
Page 196: ...194 10 2 Hood Section Fig 2...
Page 198: ...196 10 3 Main Body Fig 3...
Page 202: ...200 10 5 HL Stand Fig 5...
Page 211: ......
Page 212: ...C60S220 201109...