163
Division
Product inspection
Serial No.
Section
Display/control panel
Tools and jigs to be used
No Inspection
procedure
Remarks
5
Checking the maintenance screen display (past data screen)
Turn the incubator power switch off.
While pressing the M2 switch on the rear of the display with a pointed tool,
turn the power switch on.
Check that the past data screen appears following the start screen.
Check that when the Brightness Control switch is pressed, the page number
displayed in the middle of the screen is changed, and that the graph shows
the previous data.
Check that when the Screen Changeover switch is pressed, the graph on the
lower part of the screen is changed to the SpO
2
/pulse rate graph.
6
Checking the display output
Turn the incubator power switch off.
Connect the display/control panel to the display output port on the opposite
side of the main body.
Turn the power switch on, and check that the start screen and the numerical
value screen appear.
When the Brighness Control switch is pressed, the
page number displayed is changed.
When the Screen
Changeover switch is
pressed, the graph is
changed to the
SpO
2
/pulse rate graph.
Summary of Contents for V-2200
Page 1: ......
Page 2: ......
Page 4: ......
Page 14: ...12 4 2 Display Control Panel...
Page 187: ...185 Available only for units destined for the US...
Page 189: ...187 C60F0263 Available only for units with serial No 1680000 or higher...
Page 190: ...188 C60F1003 Available only for units destined for the US...
Page 191: ...189 9 2 HL stand...
Page 192: ...190...
Page 193: ...191 10 Replacement Parts List Replacement Parts List...
Page 194: ...192 10 1 Pole Arm Upper Part of the Main Body Accessories and Humidity Chamber Section Fig 1...
Page 196: ...194 10 2 Hood Section Fig 2...
Page 198: ...196 10 3 Main Body Fig 3...
Page 202: ...200 10 5 HL Stand Fig 5...
Page 211: ......
Page 212: ...C60S220 201109...