6-32 | Model 933S UV Analyzer
f. To assign an input to a channel click the first (top) button under
Channels
to select it.
g. Next, click
Device
under
Data Source
to enable the
Input Channel
drop-down list box. From this list, select “PMT Ref1” to assign this
signal to the first channel.
Repeat Step f and Step g to assign “PMT Ref 2”, “PMT Ref 3”,
“PMT Ref 4”, “PMT Ref 5”, and “PMT Ref 6” to subsequent chan-
nels.
h. Click
OK
when finished.
11. Click
on the
Standard Toolbar
to start scanning the
Measure
and
Reference
PMT signals.
• Observe PMT signals of filter positions 1, 3, and 5 if Lamp 1 was
replaced.
• Observe PMT signals of filter positions 2, 4, and 6 if Lamp 2 was
replaced.
To change the Y-axis voltage scale, right-click over the graph. Click
Display
and change the
Upper
and
Lower Limits
.
If the lamp pulse current-control value for a filter location is negative,
that filter location is not used.
12. Align the Source Lamps (see “Source Lamp Alignment” in this chapter)
to ensure optimal operation and then perform another
Auto-Setup
.
13. Swing the
Microcontroller/Micro-Interface
board assembly back to its
original position and tighten the (2) M3 X 8 screws on the mounting
bracket.
14. Change the
Flow Control
to ‘
3
’ (
zero
). Click
Apply
(next to
Flow Con-
trol
) to set the analyzer into
zero
Flow Control mode and then click
OK
. Do not save the changes to EEPROM. Allow the Zero gas to flow
for 2–3 minutes.
NOTE
Summary of Contents for 933S
Page 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Page 34: ...2 10 Model 933S UV Analyzer...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 933S UV Analyzer...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Page 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Page 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Page 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Page 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Page 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Page 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Page 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Page 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Page 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Page 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Page 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Page 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Page 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Page 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...