Maintenance and Troubleshooting
|
6-65
Decimal Alarm
Bit
Value
Type
Alarm Condition / Description and Suggested Corrective Action
6
64
Warning
w: Sample gas pressure too low
The sample gas pressure is below 1 % of the full-scale range.
Note: This alarm is used only by analyzers equipped with the optional
pressure transducer.
This alarm can be factory-set as a Warning or a Fault, based on
customer requirements. A Warning indicates the data is suspect (the
analyzer requires service). For a description of this alarm if configured
as a Fault, refer to the “
f: Sample gas pressure too low
” alarm
description later in this section.
Corrective Action:
•
Check the sample gas pressure gauge and ensure that it is at an appropriate
flow and pressure, as recommended on system drawings.
•
From the
Output Setup
dialog box, confirm the sample gas
Low Scale
and
High Scale
pressure settings are correct.
•
Check for leaks in the sample system (see “Sample System Leak Check” in
Chapter 3).
•
Check the solenoids (backpan).
If this doesn’t correct the alarm, take appropriate safety precautions, open the
Lower Enclosure, and:
•
Check the pressure transducer.
•
Inspect the Check Valves in the Column Block Assembly for signs of wear (see
“Column Block Maintenance” in this chapter).
•
Replace the
Customer I/O
board.
Contact AMETEK to verify operation
before removing this board.
7
128
Warning
w: Low flow detected
A low sample gas flow has been detected. This alarm will be delayed by the time
(seconds) entered in the
Flow Switch Alarm
Time Delay
field on the
Alarm
Setting and Stream Switch Parameters
dialog box.
Note: This alarm is used only by analyzers equipped with the optional flow
switch.
This alarm can be factory-set as a Warning or a Fault from the
Alarm
Type
field in the
Alarm Setting and Stream Switch Parameters
dialog box, based on customer requirements. A Warning indicates the
data is suspect (the analyzer requires service). For a description of this
alarm if configured as a Fault, refer to the “
f: Low flow detected
” alarm
description later in this section.
Corrective Action:
•
Check the Probe valves to ensure they are open.
•
Check the sample system to ensure there is adequate flow and there are no
plugs.
•
If your system uses an aspirator to move the sample gas, adjust the aspirator
air to ensure a proper flow rate, as indicated on analyzer drawings.
Also, check the aspirator air line for damage or bends.
•
Check the flow switch for plugs and/or for an open circuit.
Take appropriate safety precautions, open the Upper Enclosure, and:
•
Check the
Micro-Interface
board by simulating the low flow switch output.
Setup (tab)
Alarm/
Stream Switch
Setup (tab)
Output
Summary of Contents for 933S
Page 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Page 34: ...2 10 Model 933S UV Analyzer...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 933S UV Analyzer...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Page 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Page 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Page 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Page 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Page 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Page 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Page 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Page 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Page 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Page 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Page 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Page 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Page 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Page 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...