Specifications | 2-1
SPECIFICATIONS
Methodology
Non-dispersive ultraviolet analysis for measurement of hydrogen sulfide
(H
2
S), carbonyl sulfide (COS), and methyl mercaptan (MeSH) in sales
quality (pipeline quality) natural gas. Frontal elution chromatography is
used to remove interfering species from the sample gas.
Full-Scale Ranges
The standard customer range is defined as full-scale ranges that fall with-
in the minimum and maximum allowable for a given Measuring Cell. The
specifications apply only to these ranges, are typical, and in some cases
application-specific. Calculations of minimum and maximum are based on
atmospheric pressure.
Standard Range Measuring Cells (40 cm or Smaller)
Maximum ranges are based on Measuring Cell length (expressed in cm).
H
2
S Standard Range
:
4,000 / (Cell length) Minimum ranges are 1/4 of
maximum ranges.
COS Standard Range
:
20,000 / (Cell length) Minimum ranges are 1/5 of
maximum ranges.
MeSH Standard Range
:
10,000 / (Cell length) Minimum ranges are 1/5 of
maximum ranges.
Low Range Measuring (White) Cell Option (81.2 cm)
H
2
S
:
0–5 PPM minimum and 0–50 PPM maximum.
COS
:
0–25 PPM minimum and 0–250 PPM maximum.
MeSH
:
0–15 PPM minimum and 0–100 PPM maximum.
Summary of Contents for 933S
Page 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Page 34: ...2 10 Model 933S UV Analyzer...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 933S UV Analyzer...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Page 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Page 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Page 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Page 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Page 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Page 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Page 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Page 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Page 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Page 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Page 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Page 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Page 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Page 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...