4-2 | Model 933S UV Analyzer
System 200 Configurator Software Overview – Getting Started
The System 200 Configurator Software provides a user interface to config-
ure the analyzer’s operating parameters, control analyzer functions, and
view analyzer data. Running under Windows 7 Professional Edition and
later, it provides powerful features and flexible operating modes. The soft-
ware is also necessary while performing maintenance and troubleshoot-
ing the analyzer. The majority of these functions are performed from the
93XS Properties
dialog box (Figure 4-1).
Navigating in the Software
While working with the software, the following rules apply:
• From the
Setup
tab (Figure 4-7) navigate to the various configuration
screens by pressing the corresponding
Modify
(Setup) button. For
information about saving changes to these screens, see “Working From
the Setup Tab” in this chapter.
If a button is greyed out, the function is not available for the analyzer.
If all buttons are greyed out, the software is not communicating with
the analyzer.
• To view live concentration values or information about the Optical
Bench or analyzer, choose the
Status
,
Analyzer Data
, or
Optical Bench
Data
tab from the
93XS Properties
dialog box.
• To return to the
Setup
tab from any screen, click
OK
or
Cancel
.
• Live values on the screens are updated at one-second intervals. When
switching to a different screen, it may take up to one second for a new
set of values to be displayed.
• While working with the software, use the following navigation buttons:
Press this button to view the Devices & Log window.
Press this butto n to start data scanning. See “Replacing the Source
Lamps” in Chapter 6.
Press this butto n to stop data scanning.
Press this butto n to navigate to the computer’s hard drive to locate
files, such as the Configurator file (“93XS.dyc”).
Press this button to print scanned data (such as Measure/Reference
PMT signals) from the analyzer.
Summary of Contents for 933S
Page 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Page 34: ...2 10 Model 933S UV Analyzer...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 933S UV Analyzer...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Page 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Page 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Page 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Page 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Page 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Page 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Page 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Page 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Page 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Page 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Page 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Page 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Page 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Page 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...