6-16 | Model 933S UV Analyzer
c.
Standard Range
Align the two larger-diameter holes in the Window Cover with the
two larger-diameter holes in the Measuring Cell and connect it to
the Window Cover with the (3) M4 x 35 screws. Tighten the screws
evenly.
Low Range
Hold the Measuring Cell with the sample fitting near the Aperture
Plate upward and connect it to the Window Cover with the (3)
M4 x 35 screws. Tighten the screws evenly.
9. Reconnect the sample tubing to the Measuring Cell.
10. Power up the analyzer and computer.
11. Allow the analyzer to warm up then perform a leak check on the
sample system fittings that were disconnected. See “Sample System
Leak Check” in Chapter 3.
Analyzers without Heated Cells:
If the leak check passes, continue with Step 12.
Analyzers with Heated Cells:
If the leak check passes, replace the Insulating Jacket or close the Oven
door to allow the Cell to warm up to normal operating temperature
and stabilize and then continue with Step 12.
12. Perform an Auto-Setup.
13. After the Auto-Setup is complete, change the
Flow Control
to ‘
3
’
(
zero
) and perform a manual Zero. Click
OK
then
Apply
. Do not save
the changes to EEPROM.
14. After the Zero is complete, change the
Flow Control
to ‘
0
’ (
auto
). Click
OK
then
Apply
. Save the changes to EEPROM.
15. If the analyzer uses
Measured
Pressure Compensation (optional),
return the setting to
Measured
mode. To do this, view the
Cell/Misc
Parameters
dialog box and enter ‘
0
’ in the
Default Cell Pres
(pressure
compensation) field.
Setup (tab)
Optical
Bench
<<Auto-Setup>>
Summary of Contents for 933S
Page 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Page 34: ...2 10 Model 933S UV Analyzer...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 933S UV Analyzer...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Page 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Page 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Page 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Page 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Page 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Page 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Page 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Page 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Page 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Page 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Page 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Page 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Page 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Page 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...