4-10 | Model 933S UV Analyzer
Save Configuration
Saves the current analyzer configuration parameters to a file that
can be used later to restore analyzer settings. See “Saving Analyzer
Configuration Parameters” in this chapter.
Restore Configuration
Retrieves previously saved analyzer configuration parameters
from a “.dev” file on the computer. See “Restoring Analyzer Con-
figuration Parameters” in this chapter.
Device Enabled
Checked
(default)
Allows communication with the analyzer,
and allows you to set up system configura-
tions.
Not Checked
Device communication is disabled.
Live Data
Checked
(default)
Software uses live data from the analyzer.
Not Checked
Software uses demonstration data.
Status
Displays the status of the analyzer (device) you are communicating
with. See “Configurator Software Troubleshooting” in this chapter for
information about how to correct the problem.
“OFFLINE”
The PC is not communicating with the ana-
lyzer (
Device Enabled
and/or
Live Data
not
selected).
“Online”
The PC is communicating with the analyzer
(
Device Enabled
and
Live Data
selected).
“DISABLED”
The PC is connected to the analyzer but is
not communicating with it (
Device Enabled
and
Live Data
not selected).
“Device Not Found”
The analyzer’s serial port configurations do
not match the computer’s.
“DEMO Online”
The PC is not connected to the analyzer (or
Device Enabled
is selected and
Live Data
is
not selected).
Configurator Software version
Displays the version of the Configurator Software (below
Status
).
Summary of Contents for 933S
Page 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Page 34: ...2 10 Model 933S UV Analyzer...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 933S UV Analyzer...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Page 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Page 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Page 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Page 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Page 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Page 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Page 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Page 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Page 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Page 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Page 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Page 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Page 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Page 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...