4-66 | Model 933S UV Analyzer
Wheel Speed (rpm)
The current Optical Bench Chopper (Filter) Wheel speed in revolu-
tions per minute (RPM).
Remote AZero
The current status of the remote, dry (potential free) contact, either
OFF
(closed) or
ON
(open). When
start
is displayed, the remote auto-
zero function has been initiated from a remote control room, such as a
Distributed Control System.
Warning Relay
The current status of the Warning Status
relay. When the analyzer is
operating in its
Normal
Run state, this relay is energized. This relay
is de-energized if a
Warning
alarm condition exists within the ana-
lyzer’s built-in diagnostic system (the analyzer requires attention). The
Warning contacts will also become de-energized if you manually force
the analyzer’s Flow Control Mode to any mode other than Analyzer
(
auto
) Control. When not in the
Normal state
, this field has a Red back-
ground.
Resetting of the contacts is delayed by the
Sample Delay Timer
dura-
tion; otherwise, this relay’s contacts reset automatically upon correc-
tion of the
Warning
alarm or upon switching back to Analyzer Control
Mode.
Fault Relay
When the analyzer is operating in its
Normal
Run state, this relay is
energized. This relay is de-energized if a
Fault
alarm condition exists
within the analyzer’s built-in diagnostic system (the analyzer requires
service). When not in the
Normal state
, this field has a Red back-
ground.
When a
Fault
alarm condition exists (relay is de-energized), the sample
system Flow Control mode will automatically switch to “
zero
” mode
if the analyzer’s Flow Control Mode is set to “
auto
” (Analyzer Control
Mode).
If the relay is configured to latch (default) it will not clear unless you
manually reset the alarm; if not, it will reset automatically upon cor-
rection of the
Fault
alarm.
Whenever a
Fault
alarm condition exists, the text on the
Display
mes-
sage lines on the
Optical Bench
,
Sensors
,
Gas Calibration
,
and
Infor-
mation
dialog boxes will turn Red to indicate a fault status.
Summary of Contents for 933S
Page 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Page 34: ...2 10 Model 933S UV Analyzer...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 933S UV Analyzer...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Page 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Page 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Page 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Page 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Page 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Page 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Page 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Page 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Page 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Page 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Page 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Page 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Page 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Page 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...