Maintenance and Troubleshooting
|
6-67
Decimal Alarm
Bit
Value
Type
Alarm Condition / Description and Suggested Corrective Action
2
4
Fault
f: Sample gas pressure too low
The sample pressure is below 1 % of the full-scale range.
Note: This alarm is used only by analyzers equipped with the optional
pressure transducer.
Corrective Action:
•
Check the sample gas pressure gauge and ensure that it is at an appropriate
flow and pressure
(80 PSIG) or as recommended on system drawings.
•
From the
Output Setup
dialog box and confirm the sample gas
Low Scale
and
High Scale
pressure settings are correct.
•
Check for leaks in the sample system (see “Sample System Leak Check” in
Chapter 3).
•
Check the solenoids (backpan).
If this doesn’t correct the alarm, take appropriate safety precautions, open the
Electronics Enclosure and:
•
Check the pressure transducer.
•
Inspect the Check Valves in the Column Block Assembly for signs of wear (see
“Column Block Maintenance” in this chapter).
•
Replace the
Customer I/O
board.
Contact AMETEK to verify operation
before removing this board.
3
8
Fault
f: On-Chip A/D not responding
The
Host Controller
board’s internal analog-to-digital converter (A/D) is not
responding.
Corrective Action:
Take appropriate safety precautions, open the Lower Enclosure, and:
•
Replace the
Host Controller
board.
Contact AMETEK to verify operation
before removing this board.
Setup (tab)
Output
Summary of Contents for 933S
Page 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Page 34: ...2 10 Model 933S UV Analyzer...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 933S UV Analyzer...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Page 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Page 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Page 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Page 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Page 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Page 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Page 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Page 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Page 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Page 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Page 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Page 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Page 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Page 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...