1-4 | Model 933S UV Analyzer
About the Analyzer Sample System
The sample system is mounted to the Model 933S Analyzer backpan. The
purpose of the system is to maintain constant pressure and temperature,
and to condition the sample to obtain correct analysis.
The sample system consists of (see Figures 1-2 and 1-3):
Component Function
Sample Probe
Extends into the sample stream to obtain a representative sample.
Contains a shut-off valve used to isolate the analyzer’s sample system
from the sample stream.
Pressure Regulator
The pressure regulator adjusts the sample gas pressure to the ana-
lyzer operating pressure (typically 80 PSIG/560 KPAG) if the process
pressure is higher than the analyzer operating pressure (refer to Final
“As-Built” drawings in the analyzer Documentation Package).
Pressure Gauge
The pressure gauge is used to set the proper sample pressure in the
sample system.
Solenoid Block
The solenoid manifold block directs flow between the Zero, Column A,
and Column B solenoids. There is also a pressure relief valve to safe-
guard against over pressuring of any components down stream of the
Solenoid Block or if the pressure regulator fails. This valve is factory-
adjusted to open at approximately 150 PSIG (1050 KPAG).
Column Switching Solenoids Solenoids are automatically turned on or off to initiate switching the
sample gas between columns.
Zero Solenoid
The zero solenoid allows Zero gas (or Span gas) to flush the sample
system during the Auto-Zero (or Span) process. The column switching
solenoids are closed when the zero solenoid is open.
Column Block
The Column Block is where the sample conditioning process takes
place. The Column Block incorporates a proprietary material to permit
correct sample analysis.
Heat Exchanger Block
The Heat Exchanger Block reduces the temperature of the conditioned
gas sample prior to it entering the Measuring Cell.
Measuring Cell
The Measuring Cell is a chamber of known length, with optical-grade
quartz windows, through which the radiation beam passes through the
gas sample. The windows are sealed by means of o-rings and remov-
able end caps.
The Measuring Cell can be unheated (typical) or heated (optional) to
keep the sample gas above the dewpoint. Other variations of Measur-
ing Cells are also available.
Flow Indicator
The flow meter provides a visual indication of the sample flow rate in
liters/minute through the analyzer. Note that the scaling on the flow
indicator is for air at standard temperature and pressure. Its purpose is
to indicate and ensure that appropriate flow conditions are maintained
and not to provide an absolute measure of the flow of sample gas.
Some systems use a Rotameter instead of the Flow Indicator.
Filterblock Assembly (optional)
The Filterblock Assembly removes condensables and particulates from
the sample gas.
Summary of Contents for 933S
Page 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Page 34: ...2 10 Model 933S UV Analyzer...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 933S UV Analyzer...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Page 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Page 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Page 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Page 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Page 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Page 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Page 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Page 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Page 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Page 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Page 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Page 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Page 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Page 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...