Maintenance and Troubleshooting
|
6-61
A
Warning
indicates the analyzer may require service and the data may be
suspect. Though all Warning alarms cause the Warning Status relay to de-
energize, the Warning Status relay can also indicate one of these analyzer
operating conditions:
•
Performing a gas calibration (Auto-Zero, Zero, Span).
•
Performing an analog output calibration.
•
Performing Optical Bench setup.
•
Flow Control is not in “
auto
”.
The Warning Status relay returns to the normal energized state automati-
cally when the offending alarm condition is corrected and/or the analyzer
completes one of the above operations.
A
Fault
indicates that a failure has been detected and the analyzer data is
not valid. Under the Fault situation, the analyzer de-energizes the Fault
Status relay and switches the sample system into “
shut-in/shut-off
” mode,
if the
Flow Control
setting is ‘
0
’ (
auto
). If the Fault Status relay is config-
ured as non-latching (
Enable latching of fail alarm
on the
Alarm Setting
and Stream Switch Parameters
dialog box is unchecked), the relay returns
to its normal energized state and the sample system is switched back to
“
Sample
” mode when the offending alarm condition is corrected. How-
ever, if the Fault Status relay is configured as latching (
Enable latching of
fail alarm
on the
Alarm Setting and Stream Switch Parameters
dialog
box is checked), the operator must unlatch the relay in one of the follow-
ing ways:
•
If the Alarm Enable digital input is permanently wired to the
Enabled
state, click the
Conc. Alarm Soft Enable
On/Off
button (
Status
tab) to
temporarily
Disable
the Concentration Alarms and then click it again
to
Enable
them.
•
If the Alarm Enable digital input is wired to a customer’s Data Acquisi-
tion System, have the system toggle that line from
Enable
to
Disable
and then back to
Enable
at least one second later.
Setup (tab)
Alarm/
Stream Switch
Summary of Contents for 933S
Page 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Page 34: ...2 10 Model 933S UV Analyzer...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 933S UV Analyzer...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Page 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Page 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Page 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Page 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Page 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Page 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Page 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Page 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Page 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Page 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Page 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Page 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Page 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Page 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...