Installation and Start-Up | 3-35
k. View the
Status
tab and check for alarms in the
HC Scode
and
MC
Scode
alarm lists. Temperature-related alarms should correct them-
selves after the analyzer has warmed up to normal operating temper-
ature. If any alarms remain active, see “Start-Up Diagnostic Checklist”
in this chapter or “Troubleshooting and Diagnostics” in Chapter 6 for
more information on how to correct alarms.
After all alarms have cleared, recheck the temperature of each
zone. Compare the current temperatures to their respective set-
point values and their start-up values (recorded earlier). Normal
operating temperatures are values that are within 5 % of the
Set-
point
values.
3. After all alarms have cleared, the analyzer will automatically perform
an Auto-Zero.
4. If all of the temperature zones have reached normal operating tempera-
tures and there are still no alarm conditions or other problems, perform
a pressure leak check on the analyzer sample system to ensure there are
no leaks (see “Sample System Leak Check” in this chapter). Then, return
to this section and complete the remaining steps.
5. After all of these checks have been made and verified, close the Up-
per/Lower Enclosures and tighten all M10 screws (9 ±1 Nm (80 in-lb)).
The sample gas should remain blocked in.
Use a soft, nonabrasive cloth to gently clean the joining areas (flame-
path) of the Upper and Lower Enclosures before closing them.
Replace the parts immediately if damage or wear is apparent. Contact
AMETEK immediately if there is any doubt about the integrity of any
flamepath.
6. Perform an Auto-Setup to optimize the PMT gains of the source
lamps. Information on how to do this is located in Chapter 6.
!
WARNING
!
CAUTION
Summary of Contents for 933S
Page 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Page 34: ...2 10 Model 933S UV Analyzer...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 933S UV Analyzer...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Page 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Page 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Page 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Page 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Page 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Page 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Page 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Page 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Page 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Page 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Page 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Page 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Page 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Page 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...