4-32 | Model 933S UV Analyzer
Kp (band)
**
The
proportional
term of the PI temperature control algorithm for each
temperature zone.
Ti (sec)
**
The
integral
term of the PI temperature control algorithm for each tem-
perature zone (seconds).
Startup (%)
The duty cycle used on start-up for each temperature zone during
steady-state operation at the setpoint.
Too Cold (°C)
This is an alarm temperature setpoint for each temperature zone. If the
current temperature of the zone falls below its
Too Cold
setpoint, it will
cause the analyzer to default to an alarm state. See also the “Tempera-
ture Control” descriptions in the following pages.
Too Hot (°C)
This is an alarm temperature setpoint for each temperature zone. If the
current temperature of the zone exceeds its
Too Hot
setpoint, it will
cause the analyzer to default to an alarm state. See also the “Tempera-
ture Control” descriptions in the following pages.
Temperature
The current temperature for each temperature zone.
Duty Cycle (%)
The current duty cycle for each temperature zone during steady-state
operation at the setpoint.
Alarm Indicator
The alarm indicator box displays active temperature-related alarms. If
any alarms are active, see “Troubleshooting and Diagnostics” in Chap-
ter 6 for details about the alarm and corrective action to take to reset
the alarm.
Summary of Contents for 933S
Page 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Page 34: ...2 10 Model 933S UV Analyzer...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 933S UV Analyzer...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Page 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Page 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Page 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Page 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Page 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Page 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Page 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Page 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Page 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Page 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Page 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Page 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Page 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Page 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...