6-66 | Model 933S UV Analyzer
Decimal Alarm
Bit
Value
Type
Alarm Condition / Description and Suggested Corrective Action
0
1
Fault
f: HC-OBMC communication problem
This serious system Fault condition indicates the
Host Controller
board cannot
establish reliable communication with the
Microcontroller
board.
Corrective Action:
Take appropriate safety precautions
and cycle the analyzer AC power off for a
minimum of 15 seconds before cycling it on again.
If this doesn’t correct the alarm, take appropriate safety precautions, open the
Lower Enclosure, and:
•
From the
Info
dialog box, click the
HC Reset
button to reset the
Host
Controller
board.
If this still doesn’t correct the alarm, open the Electronics Enclosures and:
- Reset the
Host Controller
board (press
SW100
, Lower Enclosure – see
Figure 6-17).
- Reset the
Microcontroller
board (press
SW400
, Upper Enclosure – see
Figure 6-18).
•
Check the connections of the cable between the
Host Controller
(
J300
) and
Micro-Interface
(
J104
) boards. Inspect it for damage (cuts, nicks, burn marks,
etc.).
•
Replace the
Microcontroller
,
Host Controller
, and/or
Micro-Interface
boards.
Contact AMETEK to verify operation before removing any of these
boards.
1
2
Fault
f: Analytical data not received
This serious system Fault condition indicates the
Host Controller
board is not
receiving analytical data from the
Microcontroller
board.
Corrective Action:
Take appropriate safety precautions
and cycle the analyzer AC power off for a
minimum of 15 seconds before cycling it on again.
If this doesn’t correct the alarm, take appropriate safety precautions, open the
Lower Enclosure, and:
•
From the
Info
dialog box, click the
HC Reset
button to reset the
Host
Controller
board.
If this still doesn’t correct the alarm, open the Electronics Enclosures and:
- Reset the
Host Controller
board (press
SW100
, Lower Enclosure – see
Figure 6-17).
- Reset the
Microcontroller
board (press
SW400
, Upper Enclosure – see
Figure 6-18).
•
Check the connections of the cable between the
Host Controller
(
J300
) and
Micro-Interface
(
J104
) boards. Inspect it for damage (cuts, nicks, burn marks,
etc.).
•
Replace the
Microcontroller
,
Host Controller
, and/or
Micro-Interface
boards.
Contact AMETEK to verify operation before removing any of these
boards.
Setup (tab)
Analyzer
Info
Summary of Contents for 933S
Page 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Page 34: ...2 10 Model 933S UV Analyzer...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 933S UV Analyzer...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Page 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Page 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Page 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Page 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Page 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Page 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Page 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Page 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Page 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Page 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Page 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Page 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Page 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Page 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...