Installation and Start-Up | 3-41
Sample Conditioning System Optimization
Perform this optimization procedure only after the Optical Bench and
Column temperatures have reached their normal operating tempera-
tures.
Differences in sample stream composition in various applications can
affect the operation of the chromatographic columns that are used to
remove interfering species. As the quantity of ethane and heavier hydro-
carbons in the sample stream increases, the elution time of all gas species
through the Columns tends to decrease.
This change in elution time affects the analyzer’s Column
Switching
Interval
and
Hold Time
parameters. These two parameters are factory-
set using stream composition data that was provided for the application.
View the
Column Timing
dialog box (Figure 3-23) and check to ensure
that the parameters are properly set for the application. Adjust them if
necessary.
Column Switch Time Optimization
The first UV-absorbing gas species that elute through the chromato-
graphic columns are hydrogen sulfide (H
2
S), carbonyl sulfide (COS), and
methyl mercaptan (MeSH). Because the Model 933S Analyzer can mea-
sure the concentrations of all three of these species, COS and MeSH do
not interfere with the H
2
S measurement. The concentration reading of
each species is automatically corrected for the presence of the other two
compounds. Ethyl mercaptan (EtSH) is the first UV-absorbing species to
elute that interferes with the concentration measurement of the first three
species.
During normal analyzer operation, while sample gas is flowing through
one absorbing column and concentration measurements are being per-
formed, the other column is being regenerated. Periodically, the columns
are switched before interfering species can elute from the absorbing
column. The
Switching Interval
is the time Column switch timing cycle.
When the
Switching Interval
is properly adjusted, the columns switch
before EtSH elutes from the absorbing column.
!
CAUTION
Summary of Contents for 933S
Page 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Page 34: ...2 10 Model 933S UV Analyzer...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 933S UV Analyzer...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Page 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Page 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Page 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Page 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Page 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Page 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Page 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Page 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Page 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Page 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Page 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Page 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Page 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Page 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...