Installation and Start-Up | 3-9
5.
Optional:
If using the Optical Bench Purge, connect the purge line
(black tube) to the Purge Fitting on the Optical Bench.
After the Optical Bench has been installed, use a soft, nonabrasive
cloth to gently clean the joining areas (flamepath) of the Upper Enclo-
sure and its door. Close the door and replace at least one of the M10
screws while completing the installation. This will ensure the flame-
path is not inadvertently damaged.
Installing the Sample System
The analyzer is supplied with fittings for Sample, Vent, and Zero gas line
connections (Figure 3-3). The typical the Sample Inlet/Vent Outlet con-
nections are on the Filterblock. For systems without the Filterblock, the
Sample Inlet connection is at the Sample Gas Pressure Regulator, while
the Vent Outlet connection is at the Rotameter.
Ensure all appropriate fuses are removed before terminating any wir-
ing. Refer to the Fuse Legend on the inside of the analyzer’s Lower
Enclosure for locations of these fuses.
Installing the Sample Probe
If the Sample Probe is supplied by AMETEK, refer to Sample Probe draw-
ings in Appendix A for important installation information.
Installing the Sample and Vent Lines
The Sample and Vent Lines are supplied by the customer. The tubing
(Sample, 1/8”; Vent, 1/4”) must be 316 stainless steel tubing with appropri-
ate NPT connectors. The typical Vent Line configuration is to disperse the
used sample gas to atmosphere.
Refer to the analyzer Block Diagram and Plumbing Schematic drawings in
Chapter 1 or Appendix A for important installation information.
For systems without the Filterblock installed, AMETEK recommends
installing a coalescing or membrane filter upstream of the Measuring
Cell.
The inlet sample pressure should be 560 KPAG (80 PSIG) or higher.
The Vent Line should vent to atmospheric pressure or < 0.5 PSI.
!
WARNING
!
CAUTION
NOTE
NOTE
Summary of Contents for 933S
Page 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Page 34: ...2 10 Model 933S UV Analyzer...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 933S UV Analyzer...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Page 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Page 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Page 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Page 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Page 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Page 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Page 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Page 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Page 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Page 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Page 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Page 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Page 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Page 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...