14
Open the pressure-limiting valve by opening the override cap or by removing
the finger and repeating the procedure. The pressure limiting valve should now
be activated and it should be possible to hear the expiratory flow from the valve.
Squeeze and release the resuscitator a few times to ensure that air is moving
through the valve system and out of the patient valve. (7.2)
Note: As the valve plates are moving during functional test or during ventilation
a slight sound may appear. This does not compromise the functionality of
the resuscitator.
Oxygen reservoir bag
Supply a gas flow of 5 l/min to the oxygen bag. Check that the reservoir fills.
If not, check the integrity of the two valve shutters or for a torn reservoir.
Oxygen reservoir tube
Supply a gas flow of 10 l/min to the oxygen tube. Check that the oxygen flows
out at the end of the reservoir tube. If not, check for a blocked oxygen tube.
10. Service
The resuscitator requires no scheduled maintenance other than regular cleaning,
inspection and testing.
11. Accessories and Spare parts
Accessories
For further information please refer to the directions for use of
the specific accessory
- Ambu reusable PEEP valve 10 (8.1)
A000 137 000
- Ambu reusable PEEP valve 20 (8.1)
A000 213 000
- Ambu disposable Pressure Manometer (8.2)
322 003 000
- Splash guard – Only Adult resuscitator (8.4)
245 000 707
- Extension tube – 30 cm (8.5)
209 000 701
- Extension tube – 90 cm (8.5)
209 000 702
- Extension tube – 105 cm (8.5)
209 000 703
- Adapter 28 mm (8.6)
209 000 506
- Adapter 24 mm (8.7)
209 000 507
Spare parts
A
Expiratory connector
245 000 511
A
Adult patient valve with pressure
limiting valve complete
470 000 503
A
Adult patient valve complete
295 000 501
A
Valve disc, patient valve
245 000 509
A
Adult Bag
470 000 501
A, P
Inlet valve complete
304 000 508
A, P
O
2
reservoir bag valve discs
245 000 514
A, P
Inlet valve disc
245 000 509
A, P, N
Reservoir bag complete
304 000 507
A, P
Override cap
299 000 509
P
Paediatric Bag
370 000 501
P
Expiratory connector
299 000 512
Summary of Contents for Oval Plus
Page 1: ...Instructions for use Ambu Oval Plus Silicone Resuscitator...
Page 3: ...3 6 6 1 6 2 6 3 7 8 7 1 7 2 8 1 8 2 8 3 8 4 8 5 8 6 8 7...
Page 19: ...BG 19 60 80 15 l min 2 4 l min 100 2 4 l min 100 7 O EPDM 8 8 1 6 1 6 2 6 3...
Page 21: ...BG 21 8 5 134 C 273 F 134 10 30 Ambu 30 30 8 6 8 7 6 1 6 2 6 3...
Page 46: ...46 1 Ambu Oval Plus Silicone 30 10 30 kg 22 66 lbs 10 2 10...
Page 48: ...48 9 2 1 2 2 5 2 3 1 3 2 5 3 10 40 cm H2 O 4 0kPa 4 1 40 cm H2 O 4 2 6 O2 5 1 5 2 5 3 VT f...
Page 49: ...49 EL O2 60 80 15 L min 2 4 l min 100 2 4 l min 100 7 O ring PSU PP EPDM 8 8 1 6 1 6 2 6 3...
Page 51: ...51 EL 8 5 134 C 134 C 10 30 Ambu 30 30 8 6 8 7 6 1 6 2 6 3...
Page 110: ...110 1 Ambu Oval Plus Silicone 30 kg 66 lbs 10 30 kg 22 66 lbs 10kg 22 lbs 2 10 kg 22 lbs...
Page 112: ...112 9 2 1 2 2 5 2 3 1 3 2 5 3 10 kg 22 lbs 40 cm H2 O 4 0 kPa 4 1 40 cm H2 O 4 2 6...
Page 113: ...JA 113 O2 Adult 5 1 Pediatric 5 2 Infant 5 3 VT f O2 15 60 80 2 4 100 2 4 100 7 O EPDM 8 8 1...
Page 115: ...JA 115 8 5 134 C 273 F 134 10 30 Ambu 30 30 8 6 8 7 Adult 6 1 Peadiatric 6 2 Neonate 6 3...
Page 174: ...174 1 Ambu Oval Plus Silicone 30 66 10 30 kg 22 66 lbs 10 22 2 10 22...
Page 176: ...176 5 9 2 1 2 2 5 2 3 1 3 2 5 3 10 22 40 H2 O 4 0 4 1 40 H2 O 4 2...
Page 177: ...RU 177 6 O2 5 1 5 2 5 3 VT f O2 60 80 15 2 4 100 2 4 100 7 8...
Page 179: ...RU 179 X X X 0 8 4 8 5 134 C 273 F 134 10 30 Ambu 30 30 8 6...
Page 180: ...180 8 7 6 1 6 2 6 3 9 7 1 7 2 5 10 10 11...
Page 213: ...213 ZH 1 Ambu Oval Plus Silicone 30 66 lbs 10 30 22 66 lbs 10 22 lbs 2 10 22 lbs...