12
8.2 Cleaning procedure
The following cleaning method has been validated according to the AAMI TIR 12
standard. Ambu makes no warranties concerning other cleaning and sterilisation
methods than the validated one described here. The validated method specifies
cleaning by washing machine using detergents.
Disassemble the devices prior to placing in the washer and program the washer
for the following parameters. Place devices into washer to allow for drainage:
Phase
Recirculation
time (minutes)
Water
temperature
Detergent type and
concentration (if applicable)
Pre-Wash 1
02:00
Cold tap water
NA
Enzyme wash 02:00
Hot tap water
Enzymatic detergent
Wash 1
02:00
60.0 °C (140 °F)
(Set point)
Enzymatic detergent
Rinse 1
0:15
Hot tap water
(60.0 °C) (140 °F)
NA
Dry the devices appropriately.
Sterilisation by gravity steam autoclave running a full cycle at 134 degrees Celsius
for 10 minutes at a standard pressure and then dry 30 minutes.
8.3 Parts that can be cleaned and sterilised
Applicable methods
X Applicable O
Not applicable
Machine Washing
Disinfection & sterilisation
Autoclave 134°
Patient valve
X
X
Bag
X
X
Inlet valve
X
X
Extension tube
X
X
Silicone Facemask
X
X
Silicone Facemask size OA
X
0
Oxygen reservoir bag
X
X
Oxygen reservoir tube
X
0
8.4 Detergents and chemical disinfectants
Use only brands that are compatible with the resuscitator materials to avoid
reduction in the lifetime of the materials. Follow the instructions of the manufacturer
of the detergent or chemical disinfectant for dilution and exposure time.
CAUTION
Avoid using substances containing phenol to clean the product. Phenol will cause
premature wear and degradation of the materials or reduced product life span.
Promptly remove all residue of cleaning materials from the resuscitator. Residues
may cause premature wear or reduced product life span.
The following steps are generally recommended. Select proper methods for the
resuscitator parts in question according to the table above.
Summary of Contents for Oval Plus
Page 1: ...Instructions for use Ambu Oval Plus Silicone Resuscitator...
Page 3: ...3 6 6 1 6 2 6 3 7 8 7 1 7 2 8 1 8 2 8 3 8 4 8 5 8 6 8 7...
Page 19: ...BG 19 60 80 15 l min 2 4 l min 100 2 4 l min 100 7 O EPDM 8 8 1 6 1 6 2 6 3...
Page 21: ...BG 21 8 5 134 C 273 F 134 10 30 Ambu 30 30 8 6 8 7 6 1 6 2 6 3...
Page 46: ...46 1 Ambu Oval Plus Silicone 30 10 30 kg 22 66 lbs 10 2 10...
Page 48: ...48 9 2 1 2 2 5 2 3 1 3 2 5 3 10 40 cm H2 O 4 0kPa 4 1 40 cm H2 O 4 2 6 O2 5 1 5 2 5 3 VT f...
Page 49: ...49 EL O2 60 80 15 L min 2 4 l min 100 2 4 l min 100 7 O ring PSU PP EPDM 8 8 1 6 1 6 2 6 3...
Page 51: ...51 EL 8 5 134 C 134 C 10 30 Ambu 30 30 8 6 8 7 6 1 6 2 6 3...
Page 110: ...110 1 Ambu Oval Plus Silicone 30 kg 66 lbs 10 30 kg 22 66 lbs 10kg 22 lbs 2 10 kg 22 lbs...
Page 112: ...112 9 2 1 2 2 5 2 3 1 3 2 5 3 10 kg 22 lbs 40 cm H2 O 4 0 kPa 4 1 40 cm H2 O 4 2 6...
Page 113: ...JA 113 O2 Adult 5 1 Pediatric 5 2 Infant 5 3 VT f O2 15 60 80 2 4 100 2 4 100 7 O EPDM 8 8 1...
Page 115: ...JA 115 8 5 134 C 273 F 134 10 30 Ambu 30 30 8 6 8 7 Adult 6 1 Peadiatric 6 2 Neonate 6 3...
Page 174: ...174 1 Ambu Oval Plus Silicone 30 66 10 30 kg 22 66 lbs 10 22 2 10 22...
Page 176: ...176 5 9 2 1 2 2 5 2 3 1 3 2 5 3 10 22 40 H2 O 4 0 4 1 40 H2 O 4 2...
Page 177: ...RU 177 6 O2 5 1 5 2 5 3 VT f O2 60 80 15 2 4 100 2 4 100 7 8...
Page 179: ...RU 179 X X X 0 8 4 8 5 134 C 273 F 134 10 30 Ambu 30 30 8 6...
Page 180: ...180 8 7 6 1 6 2 6 3 9 7 1 7 2 5 10 10 11...
Page 213: ...213 ZH 1 Ambu Oval Plus Silicone 30 66 lbs 10 30 22 66 lbs 10 22 lbs 2 10 22 lbs...