Commonly Used Procedures 6
Agilent Nano Indenter G200 User’s Guide
6-31
Frame Stiffness and Area Function Calibration
The first step in calibrating the frame stiffness and area function is to
acquire data on fused silica over a range of displacements.
There are three methods from which the user may determine machine
compliance. These methods are:
•
Frame Stiffness and Area Function Calibration: Express Test
using
rapid indentation.
•
Frame Stiffness Calibration: ISO 14577 Method
indentation.
•
Frame Stiffness Calibration: CSM Method
using dynamic
indentation.
See
Frame Stiffness and Indenter Area Function Theory
"Theory" for more theoretical information.
General Information
Frame stiffness and area function information are stored in the same
location. To access the information, choose
Tip > Configure Tip
from
the main menu as shown in
Figure 6-45
Select
Configure Tip
from the Tip menu
Tip definitions are stored by user-defined names. The recommended
convention to use is
Serial No. Date Method
(ie TB12938 042812
Express Test) as shown in