Commonly Used Procedures 6
Agilent Nano Indenter G200 User’s Guide
6-64
16
Choose clean location on reference material and set array size of at
least ten indentation tests.
17
Click the
Next Step
button.
18
Click
Run Batch
.
Periodically, NanoSuite will display the message shown in
This is normal because the indenter has not yet touched the surface of
the sample.
Figure 6-89
NanoSuite message
Allow testing to complete.
19
When testing is complete, go to the
Review
page.
20
Right-click anywhere in the Legend pane and select
Select All Tests
from the resulting pop-up menu.
21
Delete any obviously aberrant tests, such as the aberrant test shown
in
a
Identify the test number by referring to the legend next to the
graph of results.
b
Right-click on the number of the aberrant test in the Legend pane.
c
Select
Delete Test
from the resulting pop-up menu.
d
Click
Yes
in the Test Deletion confirmation box
e
Right-click anywhere in the Legend pane and select
Select All
Tests
from the resulting pop-up menu to populate the graph.
22
Access the frame stiffness and area function information by choosing
Tip > Configure Tip
as shown in
.
Figure 6-90
Select
Configure Tip
from Tip menu