NanoSuite Explorer D
Agilent Nano Indenter G200 User’s Guide
D-57
31
In the Insert New Variable dialog box, enter the Internal Name as
RampLoadingRate
(no spaces).
32
Under the label Add To, select the
Sample
radio button.
33
Click
OK
.
34
In the General tab, set the following values:
a
Display name =
RampLoading Rate
b
Unit class =
Force Rate
c
Unit Type =
mN/s
35
In the Description tab, enter
Loading rate during the initial ramp
.
36
In the Formula tab, click the
Edit…
button.
37
In the Formula Editor dialog box, type
RampForceLimit/RampTime
(no spaces) into the text box.
38
Click the
Check
button.
39
Verify that the response is
Formula is okay
.
40
Click
OK
.
41
In the Advanced tab, set the following values:
a
Category =
Load Segment Formulas
b
When =
Pre Sample
c
Deselect all options
.
42
Select
Validate Method
from the Tools menu.
43
Verify that you receive a message box that the method is valid.
44
On the left-hand side of the screen, click on the
Test Flow
tab.
45
Scroll down and select the
GoTo – Raw Load, Loading,
LoadingRateForStrain
test segment, as shown in