Commonly Used Procedures 6
Agilent Nano Indenter G200 User’s Guide
6-45
Frame Stiffness and Area Function Calibration: Express Test
1
Load the fused silica standard into the NanoVision stage.
2
In NanoSuite, select
Open Method…
from the Method menu.
3
From the Filename drop-down box in Open Method dialog box,
select
Express Test > Express Test Tip Calibration
as shown in
Figure 6-64
Select Express Test method
4
Click
OK
.
5
Right-click anywhere in the handset area and select
Nano HandSet
from the resulting pop-up menu, as shown in