Express Test I
Agilent Nano Indenter G200 User’s Guide
H-6
7
Click the
Run
button.
Once testing is complete, the software will prompt the user to save the
file.
8
Input a file name and click
Save
.
9
To view the results, click the
Review
tab.
The 2D data is available by selecting the
Channels
tab as shown in
Figure H-4
Examples of Results from the Channels tab
The available results are:
Points Excluded
: Points rendered invalid and excluded from reported
results by the Filter Graph function (Help menu provides further
information on this subject).
Disp Into Surface (nm)
: Peak indentation displacement, averaged over
all valid indents in the test array.
Modulus (GPa)
: Young's modulus, averaged over all valid indents in
the test array.
Modulus SD (GPa)
: Standard deviation of valid Young's modulus
measurements for the test array.
Hardness (GPa)
: Hardness, averaged over all valid indents in the test
array.
Hardness SD (GPa)
: Standard deviation of valid hardness
measurements for the test array.
Stiffness (N/m)
: Contact stiffness, averaged over all valid indents in the
test array. If the contact stiffness is more than 10 % of the frame
stiffness, best results may require the determination of frame stiffness