Express Test I
Agilent Nano Indenter G200 User’s Guide
H-3
Figure H-2
Run button in Nanosuite software
6
Once testing is complete:
a
Go to the main menu
b
Select
Tip
and then
c
Select
Configure Tip
If prompted to add the current tip to your setup, click
Yes
.
7
Click
Calibrate Tip and Frame
.
The next prompt warns you that the sample will be modified and saved.
8
Click
Yes
.
9
Edit and/or verify the inputs on the right-hand side of the screen
(beneath the Calculate button) to be:
a
DCM
b
Express Test
c
The nominal lead term supplied on the tip's certificate should be
used. If this certificate is unavailable, leave the input at its default
value of
24.5
.
10
Click the
Calculate
button.
11
Click the
Save
button.
12
In the Configure Tip Parameters window, click
OK
.
Note more information about the meaning of these calibrations and
optional settings refer to