High Load C
Agilent Nano Indenter G200 User’s Guide
C-3
Figure C-2
Open High Load method
6
Install the sample tray, with the fused silica standard in the center
hole, into the instrument according to
7
Set up a batch of samples for CSM testing:
a
Select
Batch Mode
from the Mode menu.
b
Follow the instructions for
, "Quick Start Guide" using the fused silica in the center
of the sample tray. Follow the procedure exactly as written. In the
Required Inputs section, set
Maximum Load
input to
100
and
click
Next Step
.
c
Click the
Review
tab to reveal the test results.
d
Right-click in the Legend pane and select
Select All Tests
.
e
Right-click on the graph and select
Y-Axis Channel >
Modulus (GPa)
, as shown in
N O T E
The high-load mechanism automatically engages when the test
prescribes a force that is greater than what can be achieved by the
normal electromagnetic mechanism (50 gf).