Continuous Stiffness Measurement A
Agilent Nano Indenter G200 User’s Guide
A-7
Commonly Used Procedures
Setting the Displacement Range for Averaged Properties
CSM test methods return properties as a continuous function of
displacement into the test surface. These properties may be averaged
over a specific displacement range in order to report single values for
these properties. You can set this range using the following procedure:
1
In the Data pane on the Review page, highlight the
Minimum
Calculation Depth
input and edit its value to the minimum
displacement to be used in calculating hardness and modulus.
2
Highlight the
Maximum Calculation Depth
input and edit its value
to the maximum displacement to be used in calculating hardness and
modulus.
3
Click anywhere else in the Data pane to update that field.
4
Select
Recalculate Sample
from the Tools menu.