Commonly Used Procedures 6
Agilent Nano Indenter G200 User’s Guide
6-61
Frame Stiffness Calibration: CSM Method
1
Make sure that the fused silica standard is in the center sample-disk
hole in the sample tray.
2
In NanoSuite, select
Open Method…
from the Method menu.
3
From the Filename drop-down box in Open Method dialog box,
select the correct method for the installed indenter head:
a
For the XP head, choose
XP > G-Series CSM Standard
Hardness, Modulus, and Tip Cal
as shown on the top in
.
b
For the DCM head, choose
DCM > G-Series DCM CSM
Standard Hardness, Modulus, and Tip Cal
as shown on the
bottom in
Figure 6-85
Select method for CSM Standard Hardness, Modulus, and
Tip Cal for XP head (top) or DCM head (bottom)
4
Click
OK
.