4-27
Performance Tests
7. Test Port Input Frequency Response
Test Port 1 Input Frequency Response from 3 GHz to 6 GHz
39
When the analyzer finishes the calibration sweep, connect the equipment as
shown in
.
Figure 4-15. Setup for Test Port 1 Input Frequency Response
40
Press
MEAS
,
INPUT PORTS
,
A
.
41
Press
MARKER
,
MARKER 1
,
MARKER FCTN
,
MARKER MODE MENU
,
MKR SEARCH
,
SEARCH:MIN
to put marker 1 at the minimum magnitude location of the trace.
42
Press
MARKER
,
MARKER 2
,
MARKER FCTN
,
MARKER MODE MENU
,
MKR SEARCH
,
SEARCH:MAX
to position marker 2 at the maximum magnitude location of the
trace.
43
Write the marker 1 or marker 2 reading, whichever has the largest absolute
magnitude, in the “Performance Test Record.”
Power Meter Calibration on Test Port 1 from 3 GHz to 6 GHz
44
Connect the equipment as shown in
.