![Agilent Technologies 8702D Installation Manual Download Page 255](http://html.mh-extra.com/html/agilent-technologies/8702d/8702d_installation-manual_2867953255.webp)
6-89
Performance Tests – Option 011
Performance Test Records - Option 011
Test 17. Receiver Magnitude Response (Option 002 Only)
Analyzer
Frequency
Ext.
Source
Frequency
Power
Meter
Value
Input A
Value
Input B
Value
Max. Diff.
Spec
Spec.
(dB)
Meas.
Uncer.
(dB)
16 MHz
32 MHz
_______
_______
_______
_______
± 1
± 0.05
16 MHz
48 MHz
_______
_______
_______
_______
± 1
± 0.05
31 MHz
62 MHz
_______
_______
_______
_______
± 1
± 0.05
31 MHz
93 MHz
_______
_______
_______
_______
± 1
± 0.05
61 MHz
122 MHz
_______
_______
_______
_______
± 1
± 0.05
61 MHz
183 MHz
_______
_______
_______
_______
± 1
± 0.05
121 MHz
242 MHz
_______
_______
_______
_______
± 1
± 0.05
121 MHz
363 MHz
_______
_______
_______
_______
± 1
± 0.05
180 MHz
360 MHz
_______
_______
_______
_______
± 1
± 0.05
180 MHz
540 MHz
_______
_______
_______
_______
± 1
± 0.05
310 MHz
620 MHz
_______
_______
_______
_______
± 1
± 0.05
310 MHz
930 MHz
_______
_______
_______
_______
± 1
± 0.05
700 MHz
1.4 GHz
_______
_______
_______
_______
± 1
± 0.05
700 MHz
2.1 GHz
_______
_______
_______
_______
± 1
± 0.05
1 GHz
2 GHz
_______
_______
_______
_______
± 1
± 0.05
1 GHz
3 GHz
_______
_______
_______
_______
± 1
± 0.05
1.5 GHz
3 GHz
_______
_______
_______
_______
± 1
± 0.05
Option 006
2 GHz
4 GHz
_______
_______
_______
_______
± 2
± 0.05
2 GHz
6 GHz
_______
_______
_______
_______
± 2
± 0.05
3 GHz
6 GHz
_______
_______
_______
_______
± 2
± 0.05