6-65
Performance Tests – Option 011
16. Source and Receiver Harmonics (Option 002 Only)
24
Move the receiver A input cable to receiver input B.
25
Press
SYSTEM
,
HARMONIC MEAS
,
HARMONIC OFF
,
STOP
,
1.5
,
G/n
.
26
Only for Analyzers with Option 006:
Press
STOP
,
3
,
G/n
.
27
through
to measure the source and receiver worst case
2nd harmonic for the receiver input B. Press
MEAS
,
INPUT PORTS
,
B
Step 6
.
28
through
to measure the source and receiver worst case
3nd harmonic for the receiver input B.
Receiver Harmonics
29
Connect the equipment as shown in
.
Figure 6-34. Receiver Harmonics Test Setup
30
Press
SYSTEM
,
HARMONIC MEAS
,
HARMONIC OFF
,
STOP
,
1.5
,
G/n
.
31
Only for Analyzers with Option 006:
Press
STOP
,
3
,
G/n
.
32
through
to measure the receiver worst case 2nd harmonic
for the receiver input B. Press
MEAS
,
INPUT PORTS
,
B
Step 6
.
33
through
to measure the receiver worst case 3nd
harmonic for the receiver input B.
34
Press
SYSTEM
,
HARMONIC MEAS
,
HARMONIC OFF
.
35
through
to measure the receiver worst case 2nd harmonic
Summary of Contents for 8702D