6-32
Performance Tests – Option 011
9. Receiver Input Crosstalk
In Case of
Difficulty
W A R N I N G
These servicing instructions are for use by qualified personnel only.
To avoid electrical shock, do not perform any servicing unless you are
qualified to do so.
W A R N I N G
The opening of covers or removal of parts is likely to expose
dangerous voltages. Disconnect the instrument from all voltage
sources while it is being opened.
W A R N I N G
The power cord is connected to internal capacitors that may remain
live for five seconds after disconnecting the plug from its power
supply.
1
Check for loose external RF cables. Inspect all cables for signs of damage, wear,
or faulty shielding.
2
Remove the analyzer top cover. Tighten any loose SMA connector nuts on the
four semi-rigid cables located between the A4/5/6 sampler/mixer assemblies.
3
Tighten any loose screws on the A4/5/6 sampler/mixer and the A7 pulse
generator assembly covers.
4
Examine the shielding clips on the A5 sampler/mixer and the A7 pulse
generator assemblies. In addition, inspect the shielding posts on the A10 digital
IF board assembly.
5
Verify that the analyzer front panel Type-N connectors are tight. Check for
connector damage.
6
Refer to the
Agilent 8753D Option 011 Service Guide
for additional
troubleshooting information.