4-77
Performance Tests
Performance Test Records
Test 6. Test Port Input Noise Floor Level
Frequency
Range
Test Port
IF
Bandwidth
Specification
(dBm)
Calculated
Value
Measurement
Uncertainty
50 kHz – 3 GHz
Port 1
3 kHz
– 82
____________
N/A
50 kHz – 3 GHz
Port 1
10 Hz
– 102
____________
N/A
50 kHz – 3 GHz
Port 2
10 Hz
– 102
____________
N/A
50 kHz – 3 GHz
Port 2
3 kHz
– 82
____________
N/A
Option 006
3 GHz – 6 GHz
Port 2
3 kHz
– 77
____________
N/A
3 GHz – 6 GHz
Port 2
10 Hz
– 97
____________
N/A
3 GHz – 6 GHz
Port 1
10 Hz
– 97
____________
N/A
3 GHz – 6 GHz
Port 1
3 kHz
– 77
____________
N/A
Test 7. Test Port Input Frequency Response
Frequency
Range
Test Port
Specification
(dB)
Measured
Value (dB)
Measurement
Uncertainty (dB)
300 kHz – 3 GHz
Port 2
± 1
_________________
0.15
300 kHz – 3 GHz
Port 1
± 1
_________________
0.15
Option 006
3 GHz – 6 GHz
Port 1
± 2
_________________
0.18
3 GHz – 6 GHz
Port 2
± 2
_________________
0.18