6-64
Performance Tests – Option 011
16. Source and Receiver Harmonics (Option 002 Only)
13
Press
SYSTEM
,
HARMONIC MEAS
,
HARMONIC OFF
,
DISPLAY
.
14
After one sweep, press
DATA
→
MEMORY
,
DATA/MEM
to normalize the trace and
record the fundamental.
15
Press
SYSTEM
,
HARMONIC MEAS
,
HARMONIC THIRD
to prepare the analyzer for
making a 3rd harmonic measurement.
16
After one sweep, press
MARKER FCTN
,
MARKER MODE MENU
,
MKR SEARCH
,
SEARCH
MAX
.
17
Write the marker 1 value on the "Performance Test Record."
Source and Receiver Harmonics
18
Connect the equipment as shown in
.
Figure 6-33. Source and Receiver Harmonics Test Setup
19
Press
PRESET
,
MENU
,
POWER
,
6
,
x1
.
20
Press
SYSTEM
,
HARMONIC MEAS
,
HARMONIC OFF
,
STOP
,
1.5
,
G/n
.
21
Only for Analyzers with Option 006:
Press
STOP
,
3
,
G/n
.
22
through
to measure the source and receiver worst case
2nd harmonic for the receiver input A.
23
through
to measure the source and receiver worst case
3nd harmonic for the receiver input A.