3-4
Manual Verification
Step 2. Measurement Calibration
C A U T I O N
Do
not
press
PRESET
or recall another instrument state. You must use the
instrument state that you loaded during the initialization procedure.
Step 2. Measurement Calibration
1
Press
MENU
,
POWER
, –
2
,
x1
to set the Agilent 8702D test port power to –2 dBm.
2
Press
CAL
,
CAL KITS & STDS
,
SELECT CAL KIT
,
CAL KIT:7mm
,
RETURN
,
RETURN
,
CALIBRATE
MENU
,
FULL 2-PORT
.
3
Press
ISOLATION
,
OMIT ISOLATION
,
ISOLATION DONE
.
4
Press
REFLECTION
.
5
Connect the “open” end of the open/short combination (supplied in the
calibration kit) to reference test port 1, as shown in
.
6
Press
FORWARD:OPEN
.
7
When the Agilent 8702D finishes measuring the standard, connect the “short”
end of the open/short combination to reference test port 1.
8
Press
FORWARD:SHORT
.
9
When the Agilent 8702D finishes measuring the standard, connect the 50 ohm
termination (supplied in the calibration kit) to reference test port 1.
10
Press
REFLECTION
,
FORWARD:LOAD
.
11
When the Agilent 8702D finishes measuring the standard, connect the “open”
end of the open/short combination to reference test port 2, as shown in
.
12
Press
REVERSE:OPEN
.
13
When the Agilent 8702D finishes measuring the standard, connect the “short”
end of the open/short calibration standard to reference test port 2.
14
Press
REVERSE:SHORT
.
15
When the Agilent 8702D finishes measuring the standard, connect the 50 ohm
termination to reference test port 2.
16
Press
REVERSE:LOAD
.
17
When the Agilent 8702D finishes measuring the standard, press
STANDARDS
DONE
.