![Agilent Technologies 8702D Installation Manual Download Page 149](http://html.mh-extra.com/html/agilent-technologies/8702d/8702d_installation-manual_2867953149.webp)
5-5
Automated Verification – Option 011
Step 1. Initialization
Figure 5-1. System Verification Test Setup
2
While the equipment is warming up, review the connector care information in
the
Agilent 8702D User’s Guide
.
3
Insert the verification kit disk into the analyzer’s disk drive.
4
Press
PRESET
,
SAVE/RECALL
,
SELECT DISK
,
INTERNAL DISK
.
5
If you want a printout of the verification data for all the devices, press
SYSTEM
,
SERVICE MENU
,
TEST OPTIONS
,
RECORD ON
.
If you switch the record function on, you
cannot
switch it off during the verifi-
cation procedure.
6
Position the paper in the printer so that printing starts at the top of the page.
7
Press
SYSTEM
,
SERVICE
MENU
,
TESTS
,
SYS VER TESTS
,
EXECUTE TEST
.
8
When the analyzer displays,
Sys Ver Init DONE
, the initialization
procedure is complete.
Do
not
press
PRESET
or recall another instrument state. You must use the in-
strument state that you loaded during the initialization procedure.