![Agilent Technologies 8702D Installation Manual Download Page 172](http://html.mh-extra.com/html/agilent-technologies/8702d/8702d_installation-manual_2867953172.webp)
6-6
Performance Tests – Option 011
2. Source Power Range, Linearity, and Accuracy
2
Zero and calibrate the power meter. Also, set the power meter for dBm. Refer
to the power meter operating and service manual for more information on how
to do this task.
3
Press
PRESET
,
MENU
,
POWER
,
10
,
x1
.
4
Press
MENU
,
CW FREQ
,
300
,
k/m
. Set the power meter calibration factor for this
CW frequency (except if you are using an Agilent 436A).
5
Write the power meter reading in the “First Value” column on the
“Performance Test Record”.
6
and
for the other CW frequencies listed on the
“Performance Test Record”.
7
Connect the equipment as shown in
.
Figure 6-3. Path Loss Calibration Test Setup (#2)
8
Press
MENU
,
CW FREQ
,
300
,
k/m
. Set the power meter calibration factor for this
CW frequency.