![Agilent Technologies 8702D Installation Manual Download Page 54](http://html.mh-extra.com/html/agilent-technologies/8702d/8702d_installation-manual_2867953054.webp)
3-8
Manual Verification
Step 3. Device Verification
Figure 3-5. Mismatch Device Verification Setup
17
Press
TESTS
,
30
,
x1
,
EXECUTE TEST
,
CONTINUE
.
Ignore the error message.
18
Press
MENU
,
POWER
,
–
2
,
x1
,
CAL
,
CORRECTION ON
.
19
through
until you have made all the S-parameter
measurements for the verification device.
20
When the Agilent 8702D displays,
Ver Dev 3 FAIL
, proceed to the next step
only if
all
S-parameter measurements show a
PASS
status under the same
condition mentioned in
.
21
Connect the verification mismatch as shown in
. Notice that port B
of the verification mismatch is now connected to the reference test port 1.