![Agilent Technologies 8702D Installation Manual Download Page 55](http://html.mh-extra.com/html/agilent-technologies/8702d/8702d_installation-manual_2867953055.webp)
3-9
Manual Verification
In Case of Difficulty
Figure 3-6. Mismatch Device Verification Setup
22
Press
TESTS
,
31
,
x1
,
EXECUTE TEST
,
CONTINUE
.
Ignore the displayed error message.
23
Press
MENU
,
POWER
,
–
2
,
x1
,
CAL
,
CORRECTION ON
.
24
through
until you have made all the S-parameter
measurements for the verification device.
25
The test is complete when the Agilent 8702D displays,
Ver Def 4 FAIL
, and
all
S-parameter measurements show a
PASS
status under the same condition
mentioned in
.
In Case of Difficulty
1
Inspect all connections. Do
not
disconnect the cables from the Agilent 8702D
test ports. Doing so will invalidate the calibration done earlier.
2
Press
PRESET
,
SAVE/RECALL
,
SELECT DISK
,
INTERNAL MEMORY
,
RETURN
. Using the
front panel knob, highlight the full 2-port cal done earlier and press
RECALL
STATE
.
3
Repeat “Step 3. Device Verification” on page 3-5.
4
If the test still fails, do the following to check the calibration:
Summary of Contents for 8702D
Page 1: ...Agilent 8702D Lightwave Component Analyzer Installation Guide ...
Page 10: ......
Page 46: ......
Page 89: ...4 33 Performance Tests 8 Test Port Crosstalk Figure 4 19 Agilent 8702D Bottom View ...
Page 166: ......
Page 256: ......