4-68
Performance Tests
18. Test Port Output/Input Harmonics (Option 002 Analyzers with Option 006 Only)
14
Press
SCALE REF
,
AUTO SCALE
,
SCALE/DIV
,
1
,
x1
to get a better viewing of the trace.
15
Press
SYSTEM
,
HARMONIC MEAS
,
HARMONIC THIRD
.
16
After one sweep, press
SCALE REF
,
AUTO SCALE
.
17
Press
MARKER FCTN
,
MARKER MODE MENU
,
MKR SEARCH
,
SEARCH MAX
.
18
Write the marker 1 value on the “Performance Test Record.”
Port 1 Input Worst Case 2nd Harmonic
19
Connect the equipment as shown in
.
Figure 4-39. Receiver Harmonics Test Setup
20
Press
PRESET
,
MENU
,
POWER
,
8
,
x1
.
21
Press
AVG
,
IF BW
,
10
,
x1
.
22
Press
START
,
16
,
M/
µ,
STOP
,
3
,
G/n
to set the frequency range.
23
Press
MEAS
,
Trans: E/E S12 REV
,
INPUT PORTS
,
A
.
24
After one sweep, press
DISPLAY
,
DATA
→
MEMORY
,
DATA/MEM
to normalize the
trace.
25
Press
SYSTEM
,
HARMONIC MEAS
,
HARMONIC SECOND
.
26
After one sweep, press
SCALE REF
,
AUTO SCALE
to get a better viewing of the
trace.