5-15
Automated Verification – Option 011
Step 2. Measurement Calibration
MENU
,
ONE-PATH 2-PORT
.
2
Press
ISOLATION
,
OMIT ISOLATION
,
ISOLATION DONE
.
3
Press
REFLECTION
.
4
Connect the “open” end of the open/short combination (supplied in the
calibration kit) to reference test port 1, as shown in
.
Figure 5-9. Connections for Measurement Calibration Standards
5
Press
FORWARD:OPEN
. The analyzer displays the message
WAIT--MEASURING
CAL STD
.
6
When the analyzer finishes measuring the standard, connect the “short” end of
the open/short combination to reference test port 1, as shown in
Figure 5-9
.
7
Press
FORWARD:SHORT
.
8
When the analyzer displays the message
CONNECT STD THEN PRESS KEY
TO MEASURE
, connect the 50 ohm termination (supplied in the calibration kit)
to reference test port 1, as shown in
.
9
Press
FORWARD:LOAD
.