![Agilent Technologies 8702D Installation Manual Download Page 96](http://html.mh-extra.com/html/agilent-technologies/8702d/8702d_installation-manual_2867953096.webp)
4-40
Performance Tests
10. System Trace Noise (Only for Analyzers without Option 006)
Source Match (Reverse) Calibration Coefficient
69
Press
SYSTEM
,
SERVICE MENU
,
TESTS
,
39
,
x1
,
EXECUTE TEST
.
70
When the analyzer finishes the test, repeat
through
.
Reflection Tracking (Reverse) Calibration Coefficient
71
Press
SYSTEM
,
SERVICE MENU
,
TESTS
,
40
,
x1
,
EXECUTE TEST
.
72
When the analyzer finishes the test, repeat
through
.
10. System Trace Noise (Only for Analyzers without
Option 006)
Perform this test to measure the system trace noise at a designated frequency
in both the A/R and B/R ratioed measurements.
Required
Equipment
For 50
Ω
Analyzers
Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779
For 75
Ω
Analyzers
Cable, 75
Ω
, Type-N 24-inch . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-2408
Procedure
1
Connect the equipment as shown in
.
Figure 4-24. System Trace Noise Test Setup