![Agilent Technologies 8702D Installation Manual Download Page 115](http://html.mh-extra.com/html/agilent-technologies/8702d/8702d_installation-manual_2867953115.webp)
4-59
Performance Tests
16. Test Port Receiver Phase Compression
SEARCH: MIN
.
25
Write the absolute value of the marker 2 reading in the “Measured Value”
column of the “Performance Test Record.”
26
through
for the other CW frequencies listed for Port 1
in the “Performance Test Record.”
In Case of
Difficulty
W A R N I N G
These servicing instructions are for use by qualified personnel only.
To avoid electrical shock, do not perform any servicing unless you are
qualified to do so.
W A R N I N G
The opening of covers or removal of parts is likely to expose
dangerous voltages. Disconnect the instrument from all voltage
sources while it is being opened.
W A R N I N G
The power cord is connected to internal capacitors that may remain
live for five seconds after disconnecting the plug from its power
supply.
1
If the analyzer fails “Test Port 2 Magnitude Compression”:
a
Repeat this test.
b
Replace the A6 B sampler assembly if the analyzer still fails the test.
2
If the analyzer fails “Test Port 1 Magnitude Compression”:
a
Repeat this test.
b
Replace the A5 A sampler assembly if the analyzer still fails the test.
16. Test Port Receiver Phase Compression
Perform this test to verify the compression/expansion phase relationships of
the analyzer’s test port receiver samplers.
Required
Equipment
For 50
Ω
Analyzers
Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779