Test 4. Receiver Minimum R Channel Level for External Source Mode
CW Frequency
(MHz)
Frac-N VCO
Lower Limit (MHz)
Frac-N VCO
Upper Limit (MHz)
Measured Value
(MHz)
Measurement
Uncertainty
10
49.496
50.496
____________
N/A
20
37.620
38.380
____________
N/A
100
49.005
49.995
____________
N/A
1 000
36.630
37.370
____________
N/A
3 000
58.216
59.392
____________
N/A
Option 006
4 000
39.198
39.990
____________
N/A
5 000
49.000
49.990
____________
N/A
6 000
58.802
59.990
____________
N/A