![Agilent Technologies 8702D Installation Manual Download Page 191](http://html.mh-extra.com/html/agilent-technologies/8702d/8702d_installation-manual_2867953191.webp)
6-25
Performance Tests – Option 011
8. Receiver Phase Frequency Response
Figure 6-12. Phase Frequency Response Test Setup
Phase Frequency Response from 300 kHz to 3 GHz
2
Press
PRESET
,
MENU
,
POWER
,
10
,
x1
.
3
Only for Analyzers with Option 006:
Press
STOP
,
3
,
G/n
.
4
Press
FORMAT
,
PHASE
,
MARKER FCTN
,
MARKER MODE MENU
,
STATS ON
.
5
Press
SCALE REF
, .
6
,
x1
,
ELECTRICAL DELAY
. Turn the analyzer front panel knob to
vary the electrical delay until the trace is in the most linear horizontal position.
6
Press
MENU
,
TRIGGER MENU
,
SINGLE
. Write the p-p value, which appears on the
analyzer display, on the “Performance Test Record.”
7
Press
CONTINUOUS
,
MEAS
,
INPUT PORTS
,
B/R
,
FORMAT
,
PHASE
.
8
and
.
9
Press
CONTINUOUS
,
MEAS
,
INPUT PORTS
,
A/B
,
FORMAT
,
PHASE
.
10
and
.
Phase Frequency Response from 3 GHz to 6 GHz
11
Press
START
,
3
,
G/n
,
STOP
,
6
,
G/n
,
CONTINUOUS
.
Summary of Contents for 8702D
Page 1: ...Agilent 8702D Lightwave Component Analyzer Installation Guide ...
Page 10: ......
Page 46: ......
Page 89: ...4 33 Performance Tests 8 Test Port Crosstalk Figure 4 19 Agilent 8702D Bottom View ...
Page 166: ......