![Agilent Technologies 8702D Installation Manual Download Page 209](http://html.mh-extra.com/html/agilent-technologies/8702d/8702d_installation-manual_2867953209.webp)
6-43
Performance Tests – Option 011
12. Receiver Magnitude Dynamic Accuracy
18
Connect the equipment as shown in
.
Figure 6-23. Channel B Magnitude Dynamic Accuracy Test Setup
19
On the Agilent 8702D Option 011:
a
Press
MENU
,
POWER
.
b
Using the analyzer front panel knob, adjust the test port power for a reading
of 0.000 dB on the Agilent 8902A.
c
Wait for the reading to stabilize.
20
On the Agilent 8902A, press the blue shift key and
SET REF
.
21
On the Agilent 8702D Option 011: